Shop Online

Carbamate Pesticide Analysis Columns
CARBAMATE COLUMNS
 
Catalog # Description
0840250 Carbamate Column, C8, expanded resolution, 4.0 × 250 mm & 1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL
1846250 Carbamate Column, C18, high resolution/capacity, 4.6 × 250 mm & 1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL
1846150 Carbamate Column, C18, rapid analysis, 4.6 × 150 mm & 1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL
18ECG001 Guard Cartridge Holder with 3 guard cartridges
18ECG002 Guard Cartridges, 2/pk.

Pickering Laboratories carbamate columns are guaranteed to produce the separation of carbamate residues, specified by EPA and AOAC methods. For C18 columns, two water/methanol gradient conditions are available for analysis of methanolic and water samples. Expanded resolution C8 column is capable of separating as many as 23 carbamates with either water/methanol or water/acetonitrile gradients. The difference in selectivity between Methanol and Acetonitrile protocols could be used for confirming peak identification.

    

Post-column conditions for pesticide analysis:

Reagent 1: Hydrolysis reagent CB130 or CB130.2

Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of CB910

Reactor 1: 100 °C, 0.5 mL

Reactor 2: ambient. 0.1 mL

Reagents flow rate: 0.3 mL/min

Detection:

Fluorometer: λex 330 nm, λem 465 nm

The recommended gradient conditions are subject to change without notice. This may happen because of lot changes in the columns, or improvements in the overall method.

The recommended gradient for the column will always be included in the column package, and it supersedes the information in this catalog.

 

CARbAMATE COLUMN 1846250 (4.6 X 250 mm), C18, 5 µm

 

CONDITIONS FOR AQUEOUS SAMPLES

Time (Min)

WATER %

MeOH %

100

0

1

100

0

1.1

82

18

36

30

70

39

30

70

39.1 0 100
41 0 100
41.1 100 0
55 100 0

FLOW RATE: 1 mL/min,
COLUMN TEMP.: 42 ºC, INJECTION VOLUME: Up to 400 µL

   

CONDITIONS FOR METHANOLIC SAMPLES

Time (Min)

WATER %

MeOH %

85

15

1

85

15

44

25

75

44.1

0

100

49

0

100

49.1 85 15
57 85 15

FLOW RATE: 1 mL/min,
COLUMN TEMP.: 42 ºC, INJECTION VOLUME: 10 µL

1  Aldicarb sulfoxide
2  Aldicarb sulfone
3  Oxamyl
4  Methomyl
5  3-Hydroxycarbofuran
6  Aldicarb
7  Propoxur
8  Carbofuran
  9  Carbaryl
10  1-Naphthol
11  Methiocarb
12  BDMC (internal standard)

    

Carbamate column 1846150 (4.6 x 150 mm), C18, 5 µm

CONDITIONS FOR AQUEOUS SAMPLES

Time (Min)

WATER %

MeOH %

100

0

1

100

0

1.1

88

12

26

30

70

26.1

0

100

28 0 100
28.1 100 0
38 100 0

FLOW RATE: 1 mL/min, COLUMN TEMP.: 42 ºC, INJECTION VOLUME: Up to 200 µL

   

CONDITIONS FOR METHANOLIC SAMPLES

Time (Min)

WATER %

MeOH %

85

15

0.5

85

15

28.5

30

70

28.6

0

100

33.5

0

100

33.6 85 15
41 85 15

FLOW RATE: 1 mL/min, COLUMN TEMP.: 42 ºC, INJECTION VOLUME: 10 µL

1  Aldicarb sulfoxide
2  Aldicarb sulfone
3  Oxamyl
4  Methomyl
5  3-Hydroxycarbofuran
6  Aldicarb
7  Propoxur
8  Carbofuran
  9  Carbaryl
10  1-Naphthol
11  Methiocarb
12  BDMC (internal standard)

   

Carbamate column 1846150 (4.6 x 150 mm), C8, 5 µm

CONDITIONS FOR WATER/METHANOL GRADIENT

Time (Min)

WATER %

MeOH %

85

15

2

85

15

42

30

70

46

30

70

46.1

0

100

51 0 100
51.1 85 15
59 85 15

FLOW RATE: 0.8 mL/min,
COLUMN TEMP.: 50 ºC,
INJECTION VOLUME: 10 µL for methanolic and up to 400 µL for aqueous samples

   

CONDITIONS FOR WATER / ACETONITRILE GRADIENT 

Time (Min)

WATER %

ACN %

90

10

2

90

10

46

49

51

46.1

30

70

49 30 70
49.1 90 10
59 90 10

FLOW RATE: 0.8 mL/min,
COLUMN TEMP.: 50 ºC,
INJECTION VOLUME: 10 µL for methanolic and up to 400 µL for aqueous samples

1  Aldicarb sulfoxide
2  Aldicarb sulfone
3  Oxamyl
4  Methomyl
5  3-Hydroxycarbofuran
6  Aldicarb
7  Propoxur
8  Carbofuran
  9  Carbaryl
10  1-Naphthol
11  Methiocarb
12  BDMC (internal standard)

   

separation of 23 carbamates using 0840250 column (4.0 x 250 mm)

CONDITIONS FOR WATER/METHANOL GRADIENT 

Time (Min)

WATER %

MeOH %

85

15

2

85

15

42

30

70

46

30

70

46.1

0

100

51 0 100
51.1 85 15
59 85 15

FLOW RATE: 0.8 mL/min,
COLUMN TEMP.: 50 ºC,
INJECTION VOLUME: 10 µL for methanolic and up to 400 µL for aqueous samples

   

1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxy
6 Aldicarb
7 Propoxur
8 Carbofuran
   9 Carbaryl
10 Methiocarb
11 BDMC internal standard
12 Butocarboxim sulfoxide
13 Butocarboxim sulfone
14 Ethiofencarb sulfoxide
15 Ethiofencarb sulfone
16 Butocarboxim
17 Ethiofencarb
18 Thiofanox sulfoxide
19 Thiofanox sulfone
20 Methiocarb sulfoxide
21 Methiocarb sulfone
22 3-Ketocarbofuran
23 Thiofanox
  N 1-Naphthol