Pickering Laboratories
Carbamate columns are guaranteed to produce the
separation of carbamate residues, specified by EPA
and AOAC methods. For C18 columns, two
water/methanol gradient conditions are available
for analysis of methanolic and water samples.
Expanded resolution C8 column is
capable of separating as many as 23 carbamates
with either water/methanol or water/acetonitrile
gradients. The difference in selectivity between
Methanol and Acetonitrile protocols could be used
for confirming peak identification.
Post-column
conditions for pesticide analysis:
Reagent 1: Hydrolysis reagent CB130
Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in
950 mL of CB910
Reactor 1: 100 °C, 0.5 mL
Reactor 2: ambient. 0.1 mL
Reagents flow rate: 0.3 mL/min
Detection:
Fluorometer: ex
330 nm, em 465
nm
The
recommended gradient conditions are subject to
change without notice. This may happen because of
lot changes in the columns, or improvements in the
overall method. The recommended gradient for
the column will always be in the column package,
and it supercedes the information in this catalog.
|
 |
 |
|
1 |
Aldicarb
sulfoxide |
|
2 |
Aldicarb
sulfone |
|
3 |
Oxamyl |
|
4 |
Methomyl |
|
5 |
3-Hydroxycarbofuran |
|
6 |
Aldicarb |
|
7 |
Propoxur |
|
8 |
Carbofuran |
|
9 |
Carbaryl |
|
10 |
1-Naphthol |
|
11 |
Methiocarb |
|
12 |
BDMC
(internal standard) |
|
|
0 |
100 |
|
|
1 |
100 |
|
|
1.1 |
82 |
|
|
36 |
30 |
|
|
39 |
30 |
|
|
39.1 |
0 |
|
|
41 |
0 |
|
|
41.1 |
100 |
|
|
55 |
100 |
|
Flow rate: 1 mL/min, column temperature
42 °C,
injection volume up to 400 µL |
|
|
 |
|
|
1 |
Aldicarb
sulfoxide |
|
2 |
Aldicarb
sulfone |
|
3 |
Oxamyl |
|
4 |
Methomyl |
|
5 |
3-Hydroxycarbofuran |
|
6 |
Aldicarb |
|
7 |
Propoxur |
|
8 |
Carbofuran |
|
9 |
Carbaryl |
|
10 |
1-Naphthol |
|
11 |
Methiocarb |
|
12 |
BDMC
(internal standard) |
|
|
0 |
85 |
15 |
|
1 |
85 |
15 |
|
44 |
25 |
75 |
|
44.1 |
0 |
100 |
|
49 |
0 |
100 |
|
49.1 |
85 |
15 |
|
57 |
85 |
15 |
Flow rate: 1 mL/min, column temperature
42 °C,
injection volume 10 µL |
|
 |
 |
|
1 |
Aldicarb
sulfoxide |
|
2 |
Aldicarb
sulfone |
|
3 |
Oxamyl |
|
4 |
Methomyl |
|
5 |
3-Hydroxycarbofuran |
|
6 |
Aldicarb |
|
7 |
Propoxur |
|
8 |
Carbofuran |
|
9 |
Carbaryl |
|
10 |
1-Naphthol |
|
11 |
Methiocarb |
|
12 |
BDMC
(internal standard) |
|
|
0 |
|
|
|
1 |
|
|
|
1.1 |
|
|
|
26 |
|
|
|
26.1 |
|
|
|
28 |
|
|
|
28.1 |
|
|
|
38 |
|
|
Flow rate: 1 mL/min, column temperature
42 °C,
injection volume up to 250 µL |
|
|
 |
|
|
1 |
Aldicarb
sulfoxide |
|
2 |
Aldicarb
sulfone |
|
3 |
Oxamyl |
|
4 |
Methomyl |
|
5 |
3-Hydroxycarbofuran |
|
6 |
Aldicarb |
|
7 |
Propoxur |
|
8 |
Carbofuran |
|
9 |
Carbaryl |
|
10 |
1-Naphthol |
|
11 |
Methiocarb |
|
12 |
BDMC
(internal standard) |
|
|
0 |
|
|
|
0.5 |
|
|
|
28.5 |
|
|
|
28.6 |
|
|
|
33.5 |
|
|
|
33.6 |
|
|
|
41 |
|
|
Flow rate: 1 mL/min, column temperature
42 °C,
injection volume 10 µL |
|
 |
 |
|
1 |
Aldicarb
sulfoxide |
|
2 |
Aldicarb
sulfone |
|
3 |
Oxamyl |
|
4 |
Methomyl |
|
5 |
3-Hydroxycarbofuran |
|
6 |
Aldicarb |
|
7 |
Propoxur |
|
8 |
Carbofuran |
|
9 |
Carbaryl |
|
10 |
1-Naphthol |
|
11 |
Methiocarb |
|
12 |
BDMC
(internal standard) |
|
|
0 |
|
|
|
2 |
|
|
|
42 |
|
|
|
46 |
|
|
|
46.1 |
|
|
|
51 |
|
|
|
51.1 |
|
|
|
59 |
|
|
Flow rate: 0.8 mL/min, column
temperature 37 °C,
injection volume 10 µL for methanolic and
up to 400 µl for aqueous samples |
|
|
 |
|
|
1 |
Aldicarb
sulfoxide |
|
2 |
Aldicarb
sulfone |
|
3 |
Oxamyl |
|
4 |
Methomyl |
|
5 |
3-Hydroxycarbofuran |
|
6 |
Aldicarb |
|
7 |
Propoxur |
|
8 |
Carbofuran |
|
9 |
Carbaryl |
|
10 |
1-Naphthol |
|
11 |
Methiocarb |
|
12 |
BDMC
(internal standard) |
|
|
| |