Pickering Laboratories Guaranteed Chemistry

Quality Guaranteed Pickering Laboratories HPLC post-column instruments and solutions products catalog for environmental, pharmaceutical and biochemical laboratories.
     
 

 

 

Printer Friendly Version

 

 

CARBAMATE PESTICIDE ANALYSIS COLUMNS

Pickering Laboratories Carbamate columns are guaranteed to produce the separation of carbamate residues, specified by EPA and AOAC methods. For C18 columns, two water/methanol gradient conditions are available for analysis of methanolic and water samples. Expanded resolution C8 column is capable of separating as many as 23 carbamates with either water/methanol or water/acetonitrile gradients. The difference in selectivity between Methanol and Acetonitrile protocols could be used for confirming peak identification.

Post-column conditions for pesticide analysis:

Reagent 1: Hydrolysis reagent CB130

Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of CB910

Reactor 1: 100 °C, 0.5 mL

Reactor 2: ambient. 0.1 mL

Reagents flow rate: 0.3 mL/min

Detection:

Fluorometer: ex 330 nm, em 465 nm

The recommended gradient conditions are subject to change without notice. This may happen because of lot changes in the columns, or improvements in the overall method. The recommended gradient for the column will always be in the column package, and it supercedes the information in this catalog.

 CARBAMATE COLUMN 1846250 (4.6 x 250 mm), C18, 5 µm

 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

CONDITIONS FOR AQUEOUS SAMPLES

TIME

WATER

MeOH %

  0

100  

0  

  1

100  

0  

  1.1

82  

18  

  36

30  

70  

  39

30  

70  

  39.1

0  

100  

  41

0  

100  

  41.1

100  

0  

  55

100  

0  

Flow rate: 1 mL/min, column temperature 42 °C,
injection volume up to 400 µL

 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

CONDITIONS FOR METHANOLIC SAMPLES:

TIME

WATER

MeOH %

  0

85  

15  

  1

85  

15  

  44

25  

75  

  44.1

0  

100  

  49

0  

100  

  49.1

85  

15  

  57

85  

15  

Flow rate: 1 mL/min, column temperature 42 °C,
injection volume 10 µL

 CARBAMATE COLUMN 1846150 (4.6 x 150 mm), C18 , 5 µm

 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

Conditions for Aqueous Samples:

TIME

WATER

MeOH %

  0

100  

0  

  1

100  

0  

  1.1

88  

12  

  26

30  

70  

  26.1

0  

100  

  28

0  

100  

  28.1

100  

0  

  38

100  

0  

Flow rate: 1 mL/min, column temperature 42 °C,
injection volume up to 250 µL

   
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

CONDITIONS FOR METHANOLIC SAMPLES

TIME

WATER

MeOH %

  0

85  

15  

  0.5

85  

15  

  28.5

30  

70  

  28.6

0  

100  

  33.5

0  

100  

  33.6

85  

15  

  41

85  

15  

Flow rate: 1 mL/min, column temperature 42 °C,
injection volume 10 µL

CARBAMATE COLUMN 0840250 (4.0 x 250 mm), C8, 5 µm,
expanded resolution for 23+ compounds

 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

CONDITIONS FOR WATER/METHANOL GRADIENT

TIME

WATER

MeOH %

  0

85  

15  

  2

85  

15  

  42

30  

70  

  46

30  

70  

  46.1

0  

100  

  51

0  

100  

  51.1

85  

15  

  59

85  

15  

Flow rate: 0.8 mL/min, column temperature 37 °C,
injection volume 10 µL for methanolic and
up to 400 µl for aqueous samples

 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

CONDITIONS FOR WATER/METHANOL GRADIENT

TIME

WATER

ACN %

  0

90  

10  

  2

90  

10  

  46

49  

51  

  46.1

30  

70  

  49

30  

70  

  49.1

90  

10  

  59

90  

10  

Flow rate: 0.8 mL/min, column temperature 37 °C,
injection volume 10 µL for methanolic and
up to 400 µL for aqueous samples

 SEPARATION OF 23 CARBAMATES USING 0846250 COLUMN (4.6 x 250 mm)



 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 Methiocarb
11 BDMC (internal standard)
12 Butocarboxim sulfoxide
13 Butocarboxim sulfone
14 Ethiofencarb sulfoxide
15 Ethiofencarb sulfone
16 Butocarboxim
17 Ethiofencarb
18 Thiofanox sulfoxide
19 Thiofanox sulfone
20 Methiocarb sulfoxide
21 Methiocarb sulfone
22 3-Ketocarbofuran
23 Thiofanox
N 1-Naphthol

CONDITIONS FOR WATER/METHANOL GRADIENT

TIME

WATER

MeOH %

  0

85  

15  

  2

85  

15  

  42

30  

70  

  46

30  

70  

  46.1

0  

100  

  51

0  

100  

  51.1

85  

15  

  59

85  

15  

Flow rate: 0.8 mL/min, column temperature 37 °C,
injection volume 10 µL for methanolic and
up to 400 µl for aqueous samples

 CARBAMATE COLUMNS CATALOG INFORMATION

CATALOG NO.

DESCRIPTION

0840250

Carbamate Column, C8, expanded resolution, 4.0 × 250 mm &
1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL

1846250

Carbamate Column, C18, high resolution/capacity, 4.6 × 250 mm &
1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL

1846150

Carbamate Column, C18, rapid analysis, 4.6 × 150 mm &
1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL

18ECG001

Guard Cartridge Holder with 3 guard cartridges

18ECG002

Guard Cartridges, 2/pk

 

 

 

 

Home  :  Products  :  Library  :  Service & Support  :  About Us  :  Contact Us  :  Distributor List  :  News & Events  :  Sitemap