Pickering Laboratories Guaranteed Chemistry

Quality Guaranteed Pickering Laboratories HPLC post-column instruments and solutions products catalog for environmental, pharmaceutical and biochemical laboratories.
     
 

 

 

Printer Friendly Version

 

 

In order to conserve solvents and reagents, increase sensitivity, and to provide for LC-MS confirmation we developed micro columns (2.1 × 100, 150 mm) for carbamate pesticide analysis. The 5 µm packing material (C18, C8) was bonded in the same fashion as our standard columns to assure the correct selectivity and ruggedness. In support of the standard methods (US EPA 531.1 and AOAC 985.32), the special post-column derivatization system was engineered to scale thereby providing the micro post-column reactor. Since our micro columns provide superior separation, operate at low flow rate with volatile mobile phases, and have low-bleed packing material they are ideal for LC-MS analysis of carbamates.

Post-column conditions for pesticide analysis using micro columns:

Reagent 1: Hydrolysis reagent CB130

Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of CB910

Reactor 1: 100 °C, 100 µL

Reactor 2: ambient, 20 µL

Reagents flow rate: 60 µL/min

Detection:

Fluorometer: ex 330 nm, em 465 nm

The recommended gradient conditions are subject to change without notice. This may happen because of lot changes in the columns, or improvements in the overall method. The recommended gradient for the column will always be in the column package, and it supercedes the information in this catalog.

 MICRO CARBAMATE COLUMN 1821100 (2.1 x 100 mm), C18, 5 µm

 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 1-Naphthol
11 Methiocarb
12 BDMC (internal standard)

CONDITIONS FOR METHANOLIC SAMPLES

TIME

WATER

MeOH %

  0

100  

0  

  0.1

88  

12  

  15

60  

40  

  20

55.5  

44.5  

  27

20  

80  

  27.1

0  

100  

  33

0  

100  

  33.1

100  

0  

  45

100  

0  

Flow rate: 0.2 mL/min, column temperature 42 °C,
injection volume 3 µL

CONDITIONS FOR AQUEOUS SAMPLES

TIME

WATER

MeOH %

  0

100  

0  

  1

100  

0  

  1.1

88  

12  

  18

60  

40  

  23

55.5  

44.5  

  30

20  

80  

  30.1

0  

100  

  35

0  

100  

  35.1

100  

0  

  48

100  

0  

Flow rate: 0.2 mL/min, column temperature 42 °C,
injection volume up to 150 µL

 MICRO CARBAMATE COLUMN 0821150 (2.1 x 150 mm), C8, 5 µm



 
1 Aldicarb sulfoxide
2 Aldicarb sulfone
3 Oxamyl
4 Methomyl
5 3-Hydroxycarbofuran
6 Aldicarb
7 Propoxur
8 Carbofuran
9 Carbaryl
10 Methiocarb
11 BDMC (internal standard)
12 Butocarboxim sulfoxide
13 Butocarboxim sulfone
14 Ethiofencarb sulfoxide
15 Ethiofencarb sulfone
16 Butocarboxim
17 Ethiofencarb
18 Thiofanox sulfoxide
19 Thiofanox sulfone
20 Methiocarb sulfoxide
21 Methiocarb sulfone
22 3-Ketocarbofuran
23 Thiofanox
N 1-Naphthol

 

CONDITIONS FOR METHANOLIC SAMPLES

TIME

WATER

MeOH %

  0

100  

0  

  0.1

90  

10  

  23

56  

44  

  30

56.5  

44.5  

  40

25  

75  

  41

0  

100  

  45

0  

100  

  45.1

100  

0  

  58

100  

0  

Flow rate: 0.17 mL/min, column temperature 42 °C,
injection volume 3 µL

 CARBAMATE MICRO COLUMNS CATALOG INFORMATION

CATALOG NO.

DESCRIPTION

0821150

Micro Analytical Column, C8, 2.1 × 150 mm & 1700-0063 Carbamate Test Mixture, 2.5 µg/mL, 1.5 mL

1821100

Micro Analytical Column, C18, 2.1 × 100 mm & 1700-0063 Carbamate Test
Mixture, 2.5 µg/mL, 1.5 mL

1700-0204

Guard Cartridge Holder with 3 guard cartridges, for use with Micro columns

1700-0205

Guard Cartridges, 2/pk, for use with Micro columns

 

 

 

 

Home  :  Products  :  Library  :  Service & Support  :  About Us  :  Contact Us  :  Distributor List  :  News & Events  :  Sitemap