Pickering Laboratories Guaranteed Chemistry

Quality Guaranteed Pickering Laboratories HPLC post-column instruments and solutions products catalog for environmental, pharmaceutical and biochemical laboratories.
     
 
 

 

Printer Friendly Version

 

 

Pickering Laboratories cation-exchange Glyphosate column is designed and tested for analysis of Glyphosate and its primary metabolite Aminomethylphosphonic acid (AMPA). This short, isocratic method provides separation of the peaks of interest. Reproducible performance is guaranteed run to run and column to column.

Post-column conditions for Glyphosate analysis:

Reagent 1: Oxidizing reagent—100 µL of 5% Sodium Hypochlorite in 950 mL of GA116

Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of GA104

Reactor 1: 36 °C, 0.5 mL

Reactor 2: ambient. 0.1 mL

Reagents flow rate: 0.3 mL/min

Detection:

Fluorometer: ex 330 nm, em 465 nm

GLYPHOSATE COLUMN 1954150 (4.0 x 150 mm) K+ CATION-EXCHANGE,
 GUARD COLUMN (3.0 x 20 mm), K+ CATION EXCHANGE

CONDITIONS

TIME

K200 %

RG019 %

  0

100  

0  

  15

100  

0  

  15.1

0  

100  

  17

0  

100  

  17.1

100  

0  

  27

100  

0  

Flow rate: 0.4 mL/min, column temperature 55 °C,
injection volume 10 µL

 GLYPHOSATE COLUMN CATALOG INFORMATION

CATALOG NO.

DESCRIPTION

1954150

Cation-exchange Column for Glyphosate analysis, 4 × 150 mm & 1700-0080 Glyphosate test mixture, 2.5 μg/mL, 1.5 mL

1953020

Cation-exchange Guard Column, 3.0 × 20 mm

1952150

Micro Cation-exchange Column, K+ form, 2.1 × 150 mm & 1700-0080 Glyphosate test mixture, 2.5 µg/mL, 1.5 mL

1954250

Column, Glyphosate 4 x 250 mm & 1700-0080 Glyphosate test mixture,
2.5 µg/mL, 1.5 mL

 

 

 

 

 

Home  :  Products  :  Library  :  Service & Support  :  About Us  :  Contact Us  :  Distributor List  :  News & Events  :  Sitemap