Pickering Laboratories Guaranteed Chemistry

Quality Guaranteed Pickering Laboratories HPLC post-column instruments and solutions products catalog for environmental, pharmaceutical and biochemical laboratories.
     
 
 

 

Printer Friendly Version

 

 

Polyether antibiotics, such as Monensin, Narasin and Salinomycin, in raw material, premix, liquid supplements and final feeds are best quantified using HPLC with post-column derivatization. Because of selectivity of post-column reaction almost no sample cleanup is needed.

LC Conditions:

Flow Rate:

0.7 mL/min, column temperature 40 °C, injection volume 10 µL

Mobile Phase:

Methanol/5% Acetic Acid in Water (90:10), isocratic

Post-column conditions  (non-metallic post-column derivatization system is required):

Reagent 1:

Concentrated Sulfuric acid / Methanol
(4:96 v/v)

Reactor 1:

Ambient, 0.1 mL

Reagent 2:

60 g of Vanillin (or p-dimethylaminobenzaldehyde) in 950 mL
of Methanol

Reactor 2:

90 °C, 1.4 mL

Reagents Flow Rate:

0.3 mL/min

Note: Using 2-reagent system for this application greatly extends reagents life.

One-reagent method and post-column derivatization system are also available.

Detection:

UV-Vis detector: Vanillin =520 nm,
DMAB =450 nm

 

 

 POLYETHER ANTIBIOTICS CATALOG INFORMATION

CATALOG NO.

DESCRIPTION

2381750

Polyether reversed-phase column, C18, 4.6 × 250 mm

18ECG001

Guard Cartridge Holder with 3 cartridges

18ECG002

Guard Cartridges, 2/pk.

 

 

 

 

 