The stationary phase in the ALKION™ column is a rigid, non-porous polymeric phase that is surface sulfonated. The two modes of separation are ion-exchange and moderate partitioning. The low capacity of the ion-exchange resin makes it an ideal phase for the separation of strongly basic and positively charged compounds. The reversed-phase character allows for discrimination between closely related species.

The unique properties of this column allow using it for a wide range of applications.

ALKION™ COLUMN ANALYSIS OF BIOGENIC AMINES

Post-column conditions:

Reagent 1: 300 mg of OPA, 2g Thiofluor™, 3 mL of 30% Brij® 35 in 950 mL of OD104

Reactor: 45 °C, 0.15 mL

Reagent flow rate: 0.3 mL/min

Detection:

Fluorometer: ex 330 nm, em 465 nm

Conditions:

TIME

K600 %

K130 %

  0

100  

0  

  12

100  

0  

  12.1

0  

100  

  15

0  

100  

  15.1

100  

0  

  20

100  

0  

Flow rate: 0.8 mL/min, column temperature 40 °C,
injection volume: 10 µL

 ALKION™ COLUMN ANALYSIS OF BIOGENIC AMINES

Post-column conditions:

Reagent 1: 300 mg of OPA, 2g Thiofluor™, 3 mL of 30% Brij® 35 in 950 mL of OD104
Reactor: 45 °C, 0.15 mL
Reagent flow rate: 0.3 mL/min

Detection:

Fluorometer: ex 330 nm, em 465 nm

1 Histamine
2 Putrescine
3 Cavaverine
4 Spermidine
5 Spermine

 

 

Conditions:

TIME [min]

K600 [%]

K563 [%]
K130 [%]

  0

100  

0  

0  

  6

100  

0  

0  

  15

0  

100  

0  

  21

0  

100  

0  

  21.1

0  

0  

100  

  23

0  

0  

100  

  23.1

100  

0  

0  

  29

100  

0  

0  

Flow rate: 0.8 mL/min, column temperature 40 °C,
injection volume 10 µL

 ALKION™ COLUMN ANALYSIS OF PARAQUAT AND DIQUAT IN WATER

Post-column conditions:

Reagent 1: 300 mg/L of Sodium Dithionite in 0.3 M Sodium Hydroxide

Reactor: 35 °C, 0.15 mL

Reagent flow rate: 0.3 mL/min

Detection:

UV-Vis detector: =395 nm for PQ, =378 nm for DQ

PQ Paraquat
DQ Diquat

 

Note: Above chromatogram is approximately 1 ppm in 200 µL.

Conditions:

TIME

1700-1101 %

1700-1102 %

1700-1103 %

i-PROPANOL %

  0

31.5  

53.5  

0  

15  

  10

28  

47  

10  

15  

  15

28  

47  

10  

15  

  15.1

0  

85  

10  

5  

  17

0  

85  

10  

5  

  17.1

31.5  

53.5  

0  

15  

  23

31.5  

53.5  

0  

15  

Flow rate: 0.8 mL/min, column temperature 50 °C,
injection volume: up to several µL

 ALKION™ COLUMN ANALYSIS OF AMINOGLYCOSIDE ANTIBIOTICS IN FEEDS

Post-column conditions:

Reagent 1: 300 mg of OPA, 2g Thiofluor™, 3 mL 30 % Brij® 35 in 950 mL of OD104

Reactor: 45 °C, 0.15 mL

Reagent flow rate: 0.3 mL/min

Detection:

Fluorometer: ex 330 nm, em 465 nm

Conditions for Apramycin Analysis:

TIME

1700-1101 %

1700-1102 %

1700-1103 %

  0

67  

33  

0  

  5

67  

33  

0  

  15

14.7  

7.3  

78  

  20

14.7  

7.3  

78  

  20.1

0  

22  

78  

  21

0  

22  

78  

  21.1

67  

33  

0  

  28

67  

33  

0  

Flow rate: 0.8 mL/min, column temperature 40 °C,
injection volume 10 µL

Conditions for Gentamicin Analysis:

TIME

1700-1101 %

1700-1102 %

1700-1103 %

  0

43  

31  

26  

  20

9  

13  

78  

  30

9  

13  

78  

  30.1

0  

22  

78  

  31

0  

22  

78  

  31.1

43  

31  

26  

  38

43  

31  

26  

Flow rate: 0.8 mL/min, column temperature 40 °C,
injection volume 10 µL

Conditions for Neomycin Analysis:

TIME

1700-1101 %

1700-1102 %

1700-1103 %

  0

60  

40  

0  

  15

13.2  

8.8  

78  

  25

11  

11  

78  

  25.1

0  

22  

78  

  26

0  

22  

78  

  26.1

60  

40  

0  

  32

60  

40  

0  

Flow rate: 0.8 mL/min, column temperature 40 °C,
injection volume 10 µL

Conditions for separation of Apramycin, Gentamicin and Neomycin :

TIME

1700-1101 %

1700-1102 %

1700-1103 %

  0

60  

40  

0  

  15

13.2  

8.8