Method Abstract
for Post-column Liquid Chromatography 107
Here you can download this material in PDF format*
Phenylketonuria (PKU)
and Maple syrup urine disease (MSUD,
branched-chain ketoaciduria) are genetically
transmitted metabolic diseases. PKU results in
elevated Phenylalanine levels due to deficient
activity of the enzyme Phenylalanine hydroxylase.
In the MSUD state, elevated levels of leucine (Leu),
isoleucine (Ile), valine (Val), and
alloiso-leucine can be detected in blood. The
increased concentration of the branchedchain amino
acids is caused by an inherited deficient activity
of the enzyme branched-chain keto-acid
decarboxylase. Failure to diagnose and treat
these conditions in newborns can result in severe
mental retardation (PKU) or even death (MSUD).
Dietary restriction of Val, Leu, and Ile for MSUD
newborns must begin immediately once the diagnosis
is made or suspected. MSUD patients must be
monitored on a regular schedule in conjunction
with the dietary therapy because even relatively
small changes in the intake of the branched-chain
amino acids may produce rather large fluctuations
of their concentrations in the plasma. In this
respect, MSUD is more difficult to control by
dietary means than phenylketonuria (PKU), because
each of the branched-chain amino acids has to be
individually adjusted and monitored. The Pickering
Laboratories PKU/MSUD post-column system is a
rapid, automated method for the quantitative
analysis of Met, Leu + Ile, Tyr and Phe from whole
blood samples. Between-injection time is 13 min
while actual time is only 7 min. The Pickering
lithium ion-exchange column has a matched set of
eluant and regenerant to provide the best possible
separation, sensitivity, and reproducibility. The
post-column derivatizing reagent is TRIONE®
Ninhydrin.
|
Analytical
Conditions |
|
COLUMN: |
PKU/MSUD column, 4.0 x 50
mm Catalog No. 0354050 |
|
TEMPERATURE: |
65 °C |
|
FLOW RATE: |
0.4 mL/min |
|
MOBILE PHASE: |
Eluant (A): Li357,
Regenerant (B): RG003 |
|
GRADIENT: |
0–8.0 min: 100% A
8.1–10.0 min: 100% B
10.1–13.0 min: 100% A |
|
|
|
POST-COLUMN SYSTEM: |
Pinnacle PCX |
|
REACTION DWELL: |
0.5 mL |
|
TEMPERATURE: |
130 °C |
|
REAGENT: |
TRIONE® Ninhydrin |
|
FLOW RATE: |
0.3 mL/min |
|
DETECTION: |
Visible @ 570 nm |

This
chromatogram is reproduced by courtesy of Quynh
Vantu and Wanda Andrews,
Virginia Newborn Screenng Laboratory, Richmond,
Virginia.