Pickering Laboratories Guaranteed Chemistry

Quality Guaranteed Pickering Laboratories HPLC post-column instruments and solutions products catalog for environmental, pharmaceutical and biochemical laboratories.
     
 

 

 

Printer Friendly Version





 

Method Abstract for Post-column Liquid Chromatography 121

Here you can download this material in PDF format*

SAJI GEORGE, WENDY RASMUSSEN, MARIA OFITSEROVA, PH.D.,
MICHAEL PICKERING, PH. D.
PICKERING LABORATORIES, INC., MOUNTAIN VIEW, CALIFORNIA, USA

Nitrite and nitrate are added to food to preserve the color and taste and to prevent foods from becoming rancid. They are also used in food for their anti-microbial properties. Higher levels in vegetables and leafy greens are possible from the use of nitrate fertilizers and/or livestock manure. Nitrite levels in food could also be produced by reduction of nitrate to nitrite during processing.

AOAC Official method 993.031 for the analysis of nitrate involves reduction using spongy Cadmium which is toxic and carcinogenic. FDA improved on this method by using Vanadium(III) chloride and heat2 for the post-column reduction of nitrate to nitrite. Nitrite reacts with this modified Griess reagent to produce a red chromophore with maximal absorbance at 535nm.

METHOD
  

Equipment

  • LC with a binary pump

  • UV/VIS detector

  • Pickering Laboratories single reagent Pinnacle PCX post-column derivatization unit

  • Pickering Laboratories anion exchange column, 4.6 X 150mm (Cat. No. 0755150)

Reagents

Sodium acetate

Vanadium(III)chloride

N-(1-Naphthyl)ethylenediamine dihydrochloride

m-Nitro aniline

20%(v/v) Hydrochloric acid

LC Conditions

LC COLUMN TEMPERATURE:

40˚C

SAMPLE INJECTION VOLUME:

10mL

LC FLOW RATE:

1mL/min

MOBILE PHASE:

15mM Sodium acetate

Post-Column Conditions 

REACTOR VOLUME:

0.5mL

REACTOR TEMPERATURE:

100˚C

REAGENT FLOW RATE:

0.1mL/min

REAGENT:

(i)

1% Vanadium(III)chloride in 20% HCl

(ii)

1% m-Nitro aniline in 20%HCl

(iii)

1% N-(1-Naphthyl) ethylenediamine
dihydrochloride in 20% HCl

 

Mix 50mL of (i) and (ii), and 1.25mL of (iii)
and dilute to 250mL using 20%HCl

DETECTOR:

UV/VIS, λmax = 535nm

Sample Preparation

To 5g of baby food in a 50mL centrifuge tube add 25mL of 5060˚C water (for vegetables) or 15mM Sodium acetate (for fruits) and shake for 10min. Add 12.5mL of acetonitrile and make up the volume to 50mL using water (for vegetables) and sodium acetate (for fruits). Centrifuge the mixture for 15mins at 5000rpm. Filter the supernatant through a 0.45μ nylon filter and dilute to fall within the linear range.

20ppm Nitrate and Nitrite Standard

Calibration Curve for Nitrite and Nitrate

Recovery Data

Sample

Spiked Conc. (mg/kg)

Calc. Conc. (mg/kg)

Recovery (%)

 

Nitrite

Nitrate

Nitrite

Nitrate

Nitrite 

Nitrate

Sweet Potato

500

500

573

580

115

116

    

2500

2500

2714

2667

109

107

Pears      

500

500

544

550

109

110

 

2500

2500

2808

2710

112

108

Apple Sauce

500

500

565

546

113

109

 

2500 

2500

2839 

2657

114

106

 

NOTES:

Post-column reagent solutions are stored in plastic or Teflon containers. All solutions are filtered through 0.45μ nylon filter before use. Nitrate/Nitrite standards should be checked prior to use for oxidation. Sample pH should be checked to determine the choice of extraction solution since acidic pH facilitates the conversion of nitrite to nitrate.

ACKNOWLEDGMENTS:

John A. Casanova, Food and Drug Administration, 60 8th Street, Atlanta, GA 30309

 

REFERENCES:

AOAC- Official Methods of Analysis of AOAC International (2000) 17th Ed., Section 50.1.11. Use of Griess Reagents Containing Vanadium (III) for the Post-Column Derivatization and Simultaneous Determination of Nitrite and Nitrite in Baby Food. Casanova, John A.; Gross, Lois K.; McMullen, Sarah E.; Schenck, Frank, JAOAC International, 2006; 89(2), 447-451.

 

* Please use Acrobat Reader to open this file

 

 

 