Pickering Laboratories Guaranteed Chemistry

Quality Guaranteed Pickering Laboratories HPLC post-column instruments and solutions products catalog for environmental, pharmaceutical and biochemical laboratories.
     
 

 

 

Printer Friendly Version

 

 

Method Abstract for Post-column Liquid Chromatography 205

Here you can download this material in PDF format*

Aminoglycoside antibiotics are among those commonly used in animal feeds to manage intestinal microorganisms. The beneficial effects include improved growth and generally healthier animal populations. Use of antibiotics creates a demand for analytical procedures to verify concentrations in pre-mixes and feeds and in some instances for residue analysis in animal products. This note describes a simple, robust analytical method for the family of Aminoglycoside antibiotics in feeds and animal products. The sample is homogenized with a generic extraction solution and the crude soluble portion is directly injected into an HPLC ion-exchange column. The column effluent is then mixed with an OPA/Thiofluor™ reagent which forms highly fluorescent derivatives with the primary amine moieties of the antibiotics.

Call if you need to analyze other Aminoglycosides.

EXTRACTION PROCEDURE:

Take one part feed:10 parts Extraction solution (w/v) Catalog No. 1700-1118 and homogenize for five minutes. Centrifuge for 10 minutes. Three layers will form: the pellet, a supernatant emulsion and a soft layer of floating fat. Carefully lift the floating fatty layer with a spatula and discard. Transfer the emulsion to a sealable vial. Coagulate the emulsion by placing the vial in a boiling water bath for 15 minutes. Centrifuge for 10 minutes. The clear supernate is filtered (0.45um Nylon) and placed in an autosampler vial.

METHOD

Analytical Conditions

COLUMN:

ALKION™ cation-exchange, K+ form, 4 x 150 mm,
Catalog No. 9410917

ALKION™ Guard column, 3 x 20 mm, Catalog No. 9493020

TEMPERATURE:

40 °C

FLOW RATE:

0.8 mL/min

MOBILE PHASE:

K01, Potassium buffer
K02, Potassium titrant
K03, Potassium ionic strength adjuster

Post-column Conditions

POST-COLUMN  SYSTEM:

Pinnacle PCX

REACTOR VOLUME:

0.15 mL

TEMPERATURE:

45 °C

REAGENT:

o-Phthalaldehyde/Thiofluor + Brij 35®

FLOW RATE:

0.3 mL/min

DETECTION:

Fluorometer
ex: 330 nm
em: 465 nm

 

Apramycin Analysis

Sample: 11.1 nmole Apramycin on column

TIME
[min]
K01
[%]
K02
[%]
K03
[%]
  0
67  
33  
0  
  5
67  
33  
0  
  15
14.7  
7.3  
78  
  20
0  
7.3  
78  
  20.1
0  
22  
78  
  21
0  
22  
78  
  21.1
67  
33  
0  

Gentamicin Analysis

Sample: 5 x 10-6 g Gentamicin complex on column

TIME
[min]
K01
[%]
K02
[%]
K03
[%]
  0
43  
31  
26  
  20
9  
13  
78  
  30
9  
13  
78  
  30.1
0  
22  
78  
  31
0  
22  
78  
  31.1
43  
31  
26  

 

Neomycin Analysis

Sample: 2.75 nmole Neomycin on column

TIME
[min]
K01
[%]
K02
[%]
K03
[%]
  0
60  
40  
0  
  15
13.2  
8.8  
78  
  25
11  
11  
78  
  25.1
0  
22  
78  
  26
0  
22  
78  
  26.1
60  
40  
0  

“ALKION” and Thiofluor are trademarks of Pickering Laboratories, Inc.
“Brij 35” is a registered trademark of ICI Americas, Inc.

* Please use Acrobat Reader to open this file

 

 

 

Home  :  Products  :  Library  :  Service & Support  :  About Us  :  Contact Us  :  Distributor List  :  News & Events  :  Sitemap

1-800-654-3330  :  ©2007 Pickering Laboratories, Inc. -- All Rights Reserved