Overview
Glyphosate Herbicide Analysis Columns
Pickering Laboratories cation-exchange Glyphosate column is designed and tested for analysis of Glyphosate and its primary metabolite Aminomethylphosphonic acid (AMPA). This short, isocratic method provides separation of the peaks of interest. Reproducible performance is guaranteed run-to-run and column-to-column.
Post-column Conditions for Glyphosate Analysis:
Reagent 1: Oxidizing reagent—100 μL of 5 % Sodium
Hypochlorite in 950 mL of GA116
Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of GA104
Reactor 1: 36 °C, 0.5 mL
Reactor 2: ambient. 0.1 mL
Reagents flow rate: 0.3 mL/min
Detection:
Fluorometer: λex 330 nm, λem 465 nm
Conditions
Glyphosate Column 1954150 (4.0 X 150 MM) K+ CATION-EXCHANGE
Catalog Number 1954150
To Be Used with Cation-Exchange GARD™ Column Protection System 1700 – 3102
CONDITIONS
Time (Min) |
K2OO % |
RGO19 % |
---|---|---|
0 |
100 |
0 |
15 |
100 |
0 |
15.1 |
0 |
100 |
17 |
0 |
100 |
17.1 |
100 |
0 |
27 | 100 | 0 |
FLOW RATE: 0.4 mL/min,
COLUMN TEMP.: 55 ºC,
INJECTION VOLUME: 10 mL
GARD™
GARD™ Column Protection System
The new GARD™ uses a proprietary material to prevent matrix compounds from passing through (and thereby protecting the analytical column), but allows the analytes of interest to pass unimpeded through the GARD™ and onto the analytical column.
The new GARD™ significantly prolongs column life without band spreading or added pressure. We demonstrated, by means of a performance comparison for Amino Acid Analysis, that the use of a GARD™ will effectively protect the analytical column, will be more cost-effective for the laboratory, is easy to change, is universal to cation-exchange applications, and most importantly has zero band spreading.
With the new GARD™ Column protection system from Pickering Laboratories the same cation-exchange GARD™ can be used for nearly all cation-exchange applications.
Pricing & Ordering
Pricing & Ordering
| |||
Catalog # | Description | ||
---|---|---|---|
1954150 | Cation-exchange Column, K+ form, 4.0 × 150 mm (with Glyphosate Test Mixture 1700-0080) |
| |||
Catalog # | Description | ||
---|---|---|---|
1700-3102 | Cation-exchange GARD™ assembly: Holder w/ 2 replaceable GARD™ | ||
1700-3101 | Replacement Cation-exchange GARD™ (2/PK) | ||
1700-3100 | GARD™ Holder |