Overview
Glyphosate Sample Clean-up Column
Pickering SPE sample clean-up cartridges are used to treat samples prior to injection. The strong cation-exchange resin is composed of sulfonic acid functional groups attached to a styrene divinylbenzene copolymer lattice. The high selectivity and ruggedness provides a simple and universal method for removing matrix interference from the sample. The SPE sample clean-up cartridges are ideal for the cleanup of vegetables, fruits and crop samples in the analysis of Glyphosate and AMPA.
Recovery Percentages
Recoveries for Glyphosate
MATRIX |
GLYPHOSATE |
AMPA |
---|---|---|
Alfalfa |
129 % |
116 % |
Strawberry |
84 % |
82 % |
Broccoli |
97 % |
95 % |
Sample Preparation
SAMPLE PREPARATION
Extraction:
Take 25 g of homogenous sample and add enough water (after estimation of moisture content)
to make the total volume of water 125 mL. Blend and Centrifuge.
Matrix Specific Modification:
High water content – reduce sample amount to 12.5 g
High protein content – add 100 μL HCl to 20 mL of extract, shake and centrifuge.
High Fat content – perform the methylene chloride partition twice.
CLEAN-UP
Methylene Chloride Partition:
To 20 mL of aqueous extract add 15 mL methylene chloride. Shake for 2-3 min and centrifuge.
To 4.5 mL of aqueous layer add 0.5 mL acidic modifier solution. Shake and centrifuge.
SPE Clean-up:
Prepare SPE cartridge (refer to product abstract PA210)
Transfer 1 mL portions of SPE mobile phase (refer to method abstract MA206 for formulation)
and discard the effluent.
Elute analytes with 12 mL SPE mobile phase
Concentration:
Evaporate to dryness using rotary evaporator or a vacuum vortex-type evaporator or lyophilize
Re-dissolve in 2 mL of the SPE mobile phase
Pricing & Ordering
Pricing & Ordering
The Zen Cart table will be inserted here…