We would like to congratulate our grand prize winners of our last newsletter’s puzzle quiz: Helene Lachance from Shur-Gain Nutreco, Tom Schneider from Suffolk County Water Authority, Narjes Ghafoori from LA County Environmental Toxicology Lab, and Joy Gottlieb from New Mexico Department of Health Scientific Lab Division!
They have each won and will shortly be receiving:
a Polaroid Cube Action Flash Memory Digital Camcorder
(POLC3X) with HD-1080p from Target.com!!
We would like to thank all of you for your submissions!
The correct answer (click to enlarge):
Thank you!
Pickering Labs
Chromatography Quiz #19: Glyphosate Analysis High Pressure Troubleshooting!
Answer the three high pressure questions correctly and win a prize! Simply email your answer as well as your full contact information to Rebecca at rlsmith@pickeringlabs.com by May 15th, 2015 in order to win. You will receive email confirmation that your submission has been received. The answer to the puzzle and winner congratulations will be published in the next issue (to be anonymous, please notify Rebecca in submission).
Glyphosate Analysis for US EPA Method 547
HPLC running with Pinnacle PCX dual-pump system. Assume normal operating conditions (with respect to flow rates, temperatures, etc.).
Glyphosate Analysis normal pressures
Column Pressure | Reagent 1 Pressure | Reagent 2 Pressure |
90 bar | 230 psi | 130 psi |
After a month of analysis your pressures have changed. What is the most likely cause of the problem in each of the three scenarios below?
Scenario One:
Column Pressure | Reagent 1 Pressure | Reagent 2 Pressure |
100 bar | 430 psi | 330 psi |
- Pre-column GARD blocked
- Heated Reactor obstructed
- Ambient Reactor obstructed
Scenario Two:
Column Pressure | Reagent 1 Pressure | Reagent 2 Pressure |
100 bar | 430 psi | 430 psi |
- Pre-column GARD blocked
- Heated Reactor obstructed
- Ambient Reactor obstructed
Scenario Three:
Column Pressure | Reagent 1 Pressure | Reagent 2 Pressure |
180 bar | 230 psi | 130 psi |
- Pre-column GARD blocked
- Heated Reactor obstructed
- Ambient Reactor obstructed